Burette
2 embedded burette stands.
Burette volumes available: 1, 5, 10, 25, 50 ml.
Burette motor: 18,000 steps.
Complies with ISO 8655-3.
UV-protected encapsulated glass syringe.
Embedded operating procedures for burette exchange, air bubble removal
(Flush).
Rinse, Fill, Empty function.
Second burette operational during first burette refill.
Inputs/outputs
2 indicator electrode inputs.
1 reference electrode input.
Selectable polarised input from
–1 mA to 1 mA in 1 µA steps, DC or AC.
Differential input.
Temperature input.
Serial connections to titration workstation/PC and second ABU62 Biburette.
Measuring ranges / Resolution
-9 to 23 pH; 0.001 pH
±2000 mV; 0.1 mV
-10 to 100°C; 0.1°C
Languages
English, German, Danish, French, Italian, Spanish and Swedish.
Casing
Fully splashproof latene®.
Graphic 128 x 128 dot LCD (TPX®)
Dimensions (H x W x D)
380 x 230 x 450 mm (excl. tubing).
Weight
5 kg (excl. reagent bottles)
CE Marking
Complies with EMC directive 89/336/EEC
and LV directive 73/23/EEC.
cETLus certification issued by ITS/SEMKO
UL Standard 61010A-1.
CSA Standard C22 2 No. 1010.1-92.
RFID technology complies with R&TTE and FCC part 15 directives.
Power requirements
47.5 – 63 Hz
115/230 Vac +15 -18%.
Environmental conditions
5 to 40°C temperature
20 to 80% relative humidity.
Ordering information
ABU62 systems
The ABU62 Biburette includes a full set of connecting cables and sample
stand accessories.
The following detachable burette stands are available: 1 ml (B601), 5
ml (B605), 10 ml (B610), 25 ml (B625) or 50 ml (B650).
Metrology
To comply with ISO 9001 and ISO 17025 requirements, our metrology department
can supply calibration and verification certificates.
| Burette
specifications according to ISO 8655-3 |
| Burette
stand |
Nominal
volume |
Maximum permissible
systematic errors |
Maximum permissible
random errors |
| Type |
ml |
± % |
± µla |
± %b |
± µlc |
| B601 |
< 1 |
0.6 |
6.0 |
0.1 |
1 |
| B605 |
5 |
0.3 |
15 |
0.1 |
5 |
| B610 |
10 |
0.2 |
20 |
0.07 |
7 |
| B625 |
25 |
0.2 |
50 |
0.07 |
17.5 |
| B650 |
50 |
0.2 |
100 |
0.05 |
25 |
a Expressed as the deviation of the mean
of a tenfold measurement from the nominal volume or from the selected
volume (see ISO 8655-6:_,8.4).
b Expressed as the coefficient of variation of a tenfold measurement
(see ISO 8655-6:_,8.5).
c Expressed as the repeatability standard deviation of a tenfold
measurement (see ISO 8655-6:_,8.5).
|